ASABE Technical Library - Abstract
Member and Access Notice
Use of FI-IR and Miniature Fiber Optic Spectrometers in Measurements of Soluble Solids Content of Pear Juice
Published by the American Society of Agricultural and Biological Engineers, St. Joseph, Michigan www.asabe.org
Citation: 2009 Reno, Nevada, June 21 - June 24, 2009 095673.(doi:10.13031/2013.26932)Authors: Tong Sun, Fei Shen, Minglin Chen, Yibin Ying
Keywords: Visible/near infrared spectroscopy, Soluble solids content, Pear juice, Partial least squares, Stepwise multiple linear regression.
The objective of this research was to measure soluble solids content (SSC) of pear juice using FT-NIR and USB4000 spectrometer, also the results of FT-NIR and USB4000 spectra were compared. The absorption spectra of 228 pear juice samples were collected by FT-NIR spectrometer with wavelength range of 660-1100 nm and USB4000 spectrometer with wavelength range of 345-1040 nm. The partial least squares regression (PLSR) combined with four spectral pre-processing methods and stepwise multiple linear regression (SMLR) were used to develop calibration models. The results show that both FT-NIR spectra and USB4000 spectra are suitable for assessing SSC of pear juice, and FT-NIR spectra is better than USB4000 spectra to measure SSC of pear juice. The coefficients of determination (R2), root mean square error of calibration (RMSEC) and root mean square error of prediction (RMSEP) for FT-NIR spectra and USB4000 spectra are 0.97, 0.141, 0.167, and 0.85, 0.325, 0.370, respectively. SMLR using sensitive wavelengths for FT-NIR spectra also obtains acceptable results.
(Download PDF) (Export to EndNotes)