Calibration Procedure for Volumetric Flow Meters with Two-Phase Flow
Published by the American Society of Agricultural and Biological Engineers, St. Joseph, Michigan www.asabe.org
Citation: Paper number 033141, 2003 ASAE Annual Meeting . (doi: 10.13031/2013.14076) @2003Authors: Mark B. Major, Michael F. Kocher, Leonard L. Bashford
Keywords: Anhydrous ammonia, Calibration, Flow meters, Measuring system, Two phase flow
Application of anhydrous ammonia fertilizer across the width of applicators is known to be uneven.
One of the difficulties in working with anhydrous application equipment is that the anhydrous flowing through
the system is typically a two-phase (gas and liquid) fluid. Mass flow rate metering and measurement of twophase
flow is difficult, and most flow meters are rated for flow of liquids or gases, not two-phase flow.
Equipment, procedures, and analyses were developed for calibration of volumetric flow meters with two-phase
flow. The two-phase flow meter calibration system was calibrated using a liquid test fluid (water). The twophase
flow meter calibration system was then used to calibrate a volumetric flow meter with a liquid test fluid
(water).
The coefficient of determination (R2) and maximum percent resolution for the calibration of the two-phase flow
meter calibration system using a liquid test fluid (water) were 0.9667 and 5.69%, respectively. The coefficient
of determination (R2) and maximum percent resolution for the calibration of the volumetric flow meter using a
liquid test fluid (water) with the two-phase flow meter calibration system were 0.9991 and 2.31%, respectively.