Click on “Download PDF” for the PDF version or on the title for the HTML version.


If you are not an ASABE member or if your employer has not arranged for access to the full-text, Click here for options.

Development of a Graphene Paper-Based Flexible Solid-Contact Lead Ion-Selective Electrode and its Application in Water

Published by the American Society of Agricultural and Biological Engineers, St. Joseph, Michigan www.asabe.org

Citation:  Transactions of the ASABE. 62(2): 245-252. (doi: 10.13031/trans.12906) @2019
Authors:   Yao Yao, Yibin Ying, Jianfeng Ping
Keywords:   Graphene paper, Heavy metal, Ion-selective electrode, Potentiometry, Water sample.

Abstract. A graphene paper-based flexible solid-contact ion-selective electrode (SC-ISE) was developed to detect lead ion sensitively. Graphene paper obtained via a simple vacuum filtration method was used as the electrode substrate for direct coating of an ion-selective membrane. The Nernstian slope of the prepared paper-based potentiometric sensor toward lead ion detection was demonstrated as 29.4 mV per decade. A detection limit as low as 2.5 x 10-7 mol L-1 was achieved. Reversed chronopotentiometry and water layer test revealed that the graphene paper-based SC-ISE possessed excellent potential stability because of the hydrophobicity of graphene paper. Furthermore, reliable data were obtained from the detection of lead ion levels in real water samples using the graphene paper-based potentiometric sensor, which shows great potential in practical application.

(Download PDF)    (Export to EndNotes)